Laser/Microscope
- Image Field Size
10 cm × 12 cm - CCD Pixel Size
6.45 microns - Dynamic Range*
22 bit (>6 logs) - Depth of Field for Best Sample Focus
6 mm - Patented FieldBrite™ XT2 Technology
CV <3% across field - Laser Lifetime
20,000 hours of operation - 700 Channel Laser Source
Solid-state diode laser at 685 nm - 800 Channel Laser Source
Solid-state diode laser at 785 nm - Light Source
600 nm diffused light source - Detectors
Low-noise CCD. Thermoelectrically cooled. - Acquisition Times
Fluorescence (700 and 800 nm) channels: 30 s, 2 min, 10 min plus variable time feature. Chemiluminescence channel: 30 s, 2 min, 10 min, 60 min plus variable time feature. - Focusing
Automatic
* This dynamic range is obtained in a single acquisition.
Operation Specifications
- Operating Conditions
For indoor use only; operating temperature 15-35 °C and dew point no greater than 22 °C, non-condensing; maximum operating temperature may be reduced at elevations above 2000 m. - Power Requirements
Universal input range is between 100-127 VAC and 200-240 VAC (voltage fluctuations not to exceed 10% of the nominal voltage); 4 Amp maximum; at 120V, 0.3 Amp typical when idle; 50/60 Hz. Overvoltage Category II.
Size and Weight
- Dimensions (instrument only)
41.4 cm W × 47 cm D × 67.3 cm H
(16.3″ × 18.5″ × 26.5″)
Depth with imaging drawer open is 59.7 cm (23.5″) - Weight
27 kg (60 lbs)
Software Specifications
- Empiria Studio® Software
Requires a minimum of 8 GB of RAM; 64-bit versions of Windows 7 or 10; Mac OS Sierra or High Sierra - Image Studio™ Software
For operating system compatibility, see this table. - Please contact LI-COR for compatibility with other software packages.